Ozone Monitor Models 106-L, 106-M, 106-MH and 106-H
For Industrial Ozone Applications
For industrial ozone applications, 2B Tech offers theModel 106-L, 106-M, 106-MH and 106-H OzoneMonitors, where L, M, MH and H refer to Low (0-100 ppm), Medium(0-1,000 ppm)06-, Medium-High (0-10,000 ppm) and High (0-20 wt%)ozone concentrations, respectively. These instruments havedifferent optical path lengths in order to accommodate an extremelywide range of ozone concentrations, spanning more than eight ordersof magnitude. Additionally, the Model 106-H was designed with aflow-through path that can be pressurized to up to 50 psi forin-line measurements with ozone generators. All four models areavailable in either a standard enclosure or Industrialenclosure. Configurations for multi-channel sampling areavailable for Models 106-L, 106-M, and 106-MH.
The Model 106 series was designed as an "ozone monitor on a board"in which nearly all of the components are mounted directly to theprinted circuit board. Thus, the Model 106 may be purchased withoutthe enclosure for OEM applications. See: ModelOEM-106.
The Model 106-L is designed for ambient low measurements down tovery low ppb levels and has been designated by the EPA as a FederalEquivalent Method (FEM) for monitoring for compliance with the USClean Air Act: EQOA–0914–218.
Ourvideo describes the different models and configurationsfor our Model 106-Series Ozone Monitors, and could help determinewhich model is best for a particularapplication. Click here orscroll down on this page to view the video.
Specifications for Different Optical Benches
Model | Range | Resolution | Precision & Accuracy |
---|---|---|---|
106-L | 0-100 ppm | 0.0001 ppm (0.1 ppb) | Higher of 0.0015 ppm (1.5 ppb) or 2% of reading |
106-M | 0-1,000 ppm | 0.01 ppm | Higher of 0.01 ppm or 2% of reading |
106-MH | 0-10,000 ppm (0-1 vol%) | 0.1 ppm | Higher of 0.05 ppm or 2% of reading |
106-H | 0-20 wt% O2 or Air | 0.01 wt% | Higher of 0.01 wt% or 2% of reading |
General Specifications
Measurement Principle | UV Absorption at 254 nm |
Measurement Interval | 2 s |
Data Averaging Options | 2 s, 10 s, 1 min, 5 min, 1 hr |
Flow Rate | ~1 Liter/min; Up to 20 L/min for Model 106-H |
Data Storage | 32,736 lines (10 s avg. = 3.7 days; 5 min avg = 113days) |
Choice of Units | ppb, pphm, ppm, µg m-3, mg m-3 (for Models106-L and 106-M) ppm, mg m-3 (for Model 106-MH) vol%, wt% air, wt% O2, g/m3, g/Nm3 (for Model 106-H) |
Data Outputs | USB, RS232, 0-2.5 V Analog, 4-20 mA, LCDDisplay |
Power Requirements | 100-240 V AC, 50/60 Hz 11-28 V DC, nominally 500 mA at 12 V DC, 6 watt for Models 106-L,M, and MH 11-28 V DC, nominally 325 mA at 12 V DC, 3.9 watt for Model106-H |
Size (standard enclosure) | 3.6 x 7.9 x 9.4 inches (9 x 20 x 24 cm) |
Weight (standard and OEM) | 4.2 lb (1.9 kg) for Model 106-L; 2.8 lb (1.3 kg)for OEM-106-L 3.9 lb (1.8 kg) for Models 106-M, -MH; 2.5 lb (1.1 kg) forOEM-106-M, -MH 4.4 lb (2.0 kg) for Model 106-H; 2.9 lb (1.3 kg) for OEM-106-H |
Size (Industrial enclosure) | 16 x 14 x 9.3 inches (40.7 x 35.7 x 23.6 cm) H x Wx D |
Weight (Industrial enclosure) | ~14 lb (6.4 kg) depending on model and multichanneloptions |
Features
Measurement based on UV absorption
Low power consumption; can be battery operated (optionallithium-ion battery)
Internal data logger with real-time clock
2-s measurement interval
On-board microprocessor with interactive menus includes dataaveraging options of 10 s, 1 m, 5 m and 1 hr
USB and RS-232 output of time/date, O3 concentration,internal temperature and pressure
Analog output (0-2.5 V and 4-20 mA) of ozone concentration inuser selected units
Two 2-level relays for control purposes (e.g., control of ozonesource or turn warning light on and off); Four 2-level relays inIndustrial enclosure models and in the 3-channel and 6-channeloptional configurations
Long-life pump (15,000 hr) for Models 106-L, 106-M and106-MH
Bluetooth option for wireless data transmission
Optional configurations for 3-channel or 6-channel air samplingfor Models 106-L, 106-M, and 106-MH